Specifications
Basic Description Electrically Powered, Computer Controlled Ventilator Control Principle: Closed Loop Flow & Pressure Controller Ventilation Modes:
|
Assist Control (A/C): |
|
Volume Controlled A/C or Pressure Controlled A/C |
|
Synchronized Intermittent Mandatory Ventilation (SIMV): |
|
Volume Controlled SIMV or Pressure Controlled SIMV |
|
Adaptive Bi-Level |
|
Continuous Positive Airway Pressure (CPAP) |
|
Pressure Support Ventilation (PSV) |
Special Modes of Operation:
|
Adjustable Rise Time |
|
100% O2 Suction |
|
Easy Exhale |
|
Sigh Breath Preset Parameters by Patient Weight |
|
Adaptive Flow™ |
|
Adaptive Time™ |
|
Backup Apnea Ventilation |
Ventilation Performance and Controlled Parameters
| Respiratory Rate |
1 to 80 BPM |
| Tidal Volume |
50 to 2,000 ml |
| Inspiratory Pressure Limit |
5 to 80 cm H 2 O |
| Inspiratory Time |
Adaptive Time TM, or 0.3 to 3 seconds |
| Peak Flow |
Adaptive Flow TM or 1 to 120 L/min |
|
Spontaneous up to 180 L/min |
| Oxygen Mix |
(FIO2)21% to 100% |
| PEEP |
0 to 20 cm H 2 O |
| Trigger Sensitivity |
1 to 20 L/min Flow Sensitivity + off |
|
–0.5 to –20 cm H 2 O Pressure Sensitivity + off |
| PSV |
0 to 60 cm H 2 O |
| Positive Pressure Relief Valve |
80 cm H 2 O |
Monitoring and Displayed Parameters
Airway Pressure (analog bar graph & numerical)
Total Breath Rate
I:E Ratio
Exhaled Tidal Volume
Exhaled Minute Volume
Peak Flow
Inspiratory Time
Electrical Power Source (external / internal)
Battery Level
Pressure, Flow and Volume Waveforms Software Package:
Real Time Pressure and Flow Waveforms
Waveform History Browse
Trending of Monitored Data
Respiratory Diagnostics Software Package:
Pressure, Flow and Volume Loops
Lung Mechanics (Static and Dynamic C, R, MAP, RSBI(RR/Vt), Time constant)
User Adjustable Alarms
Respiratory Rate (high / low) Apnea (0 to 120 seconds)
Minute Volume (high / low) FIO2 (high / low)
Pressure (high / low) Leak (0 to 100%)
Low Vt (15%-85%)
Additional Alarms and Indicators
| Alarms: |
Indicators: |
| Inverse I:E Ratio |
Alarm Silence Icon & Timer |
| Low O2 Pressure |
Breath Type Icon |
| AC Disconnect |
Internal Battery Use |
| Low Battery |
Date and Time |
| Over Temperature |
Hour Meter |
| Service Notice |
Battery Charge Level |
| Patient Disconnect |
Need Calibration |
| Check Sensor |
LED: On, Charge, Alarm |
Size and Weight Dimensions: Height 13" / 33 cm Width 9.5" / 24 cm Depth 10.3" / 26 cm Screen 8.4" diagonal Weight 15.4 lb / 7 kg (without battery) Battery Weight 6.5 lb / 3 kg Overall Weight 22 lb / 10 kg
Power Supply
| External AC |
100 to 240 V, 50 to 60 Hz, Max 2.0 A |
| External DC |
12 to 15 V (Optional: 10 to 30 V), Max 8.5 A |
| Internal Battery |
Sealed Lead-Acid 12 V (7.8 Ah) (rechargeable) |
|
Operating Time - Up to 2 hours depending on ventilation parameters |
|
optional 4 hour battery available |
Oxygen (enrichment) Supply
| High Pressure Supply |
40 to 60 psi (2.8 to 4.2 bar) |
| Low Pressure Max |
80 L/min or 0.5 psi |
External Interface Remote Monitor (VGA) DIN Keyboard Connector RS-232 Serial Port, 9 Pin RJ11 Remote Alarm Connector
Environmental Specifications
| Operating Temperature |
0 to 50° C / 32 to 120° F |
| Storage Temperature |
?15 to 70° C / ?4 to 140° F |
| Relative Humidity |
15 to 95% at 30°C / 85°F |
| Water / Dust Resistance |
IP54 (Splash Proof) |
| Atmospheric Pressure |
430 to 825 mm Hg (15,000 feet) |
| Vibration |
IEC 68-2-6 and IEC 68-2-34 |
|
MIL-STD-810E |
| Shock |
IEC 68-2-27 (100g) |
|
MIL-STD-810E |
| Total External Sound Level |
40-45 dBa at one meter |
Standards and Safety Requirements Meets the requirements of: ASTM F1100-90 CSA C22.2 No. 601.1 / 601.2 IEC 60601-1 IEC 60601-1-2 EN 60601-2-12 EN 794-1 / 2 / 3 ISO 10651-1 / 2 / 3 UL 2601.1
|
|
|